Output Stored Data Specifiers:
Memory specifiers are TRACE, PATHCAL, and CALFACTOR
Sensor specifiers are A, B, and C
Trace specifiers are 0 through 9
Output Stored Data Format:
START FREQ,STOP FREQ (MHz) D.ddd
then 512 data points for TRACE (dB)
±
ddd.dd
then 4096 data points for PATHCAL
±
ddd.dd
then 4096 data points for CALFACTOR
±
ddd.dd
D = 0 to 6 digits
Output Stored Data Examples:
OUTPUT;TRACE 5
OUTPUT;PATHCAL A
OUTPUT;CALFACTOR C
Output Stored Data Related Command:
INPUT
Description: There are two forms of Output commands. One form outputs currently measured data.
The other form outputs stored data such as Trace memories, Path Cal memories, and
sensor Cal Factor memories. These forms have completely different syntaxes but share
the same verb. Users are cautioned to make note of the difference between the two. The
Output Current Data and the Output Stored Data descriptions starting on page 3-31.
NOTE: TRACE uses a Trace specifier, PATHCAL and CALFACTOR use Sensor specifiers.