EasyManua.ls Logo

Giga-tronics 8003 - Page 218

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8003 Precision Scalar Analyzer
4-4 Manual 20791, Rev. C, June 2001
3. Set the 432A RANGE switch to COARSE ZERO, and adjust the front panel COARSE ZERO
control to obtain a zero (
±
2% F.S.) meter indication.
4. Set the DVM to a range that results in a resolution of 1
µ
V and connect the positive and
negative input, respectively, to the V
COMP
and V
RF
connectors on the rear panel of the 432A.
5. Fine zero the 432A on the most sensitive range, then set the 432A range switch to 1 mW.
6. Record the DVM indication as V
0
.
7. Turn ON the 8003 Calibrator RF power as follows:
Press [CONFIG] [SERVICE] [SET CAL]
In the “System” section of the front panel of the 8003 is a key [SET CAL] labeled
CONFIG
.
Press this key and a menu will appear with a softkey labeled SERVICE. Press this key and
another menu will appear with a softkey labeled SET CAL. Press [SET CAL]. Record the
reading shown on the DVM as V
1
.
8. Disconnect the DVM negative lead from the V
RF
connector on the 432A, and reconnect it to
the 432A chassis ground. Record the new indication observed on the DVM as V
COMP
.
9. Press the ABORT softkey on the 8003.
10. Calculate the Calibrator Output level (P
CAL
) using the following formula:
where:
P
CAL
= calibrator output power reference level
V
COMP
= previously recorded value in Step 8
V
1
= previously recorded value in Step 7
V
0
= previously recorded value in Step 6
R= 200
(assuming HP478A-H75 mount)
Calibration factor = value for the thermistor mount at 50 MHz (traceable to the NIST)
11. Verify that the P
CAL
is within the following limits:
1 mW
±
0.019 mW (0.981 to 1.019 mW)
For record purposes, the measured value of P
CAL
can be entered on the Performance Verifica-
tion Test Data Recording Sheet starting on page 4-14.
NOTE: Ensure that the DVM input leads are isolated from chassis ground
when performing the next step.
NOTE: The V
1
reading must be taken within 15 seconds after pressing [SET
CAL]. Otherwise, press the ABORT softkey and repeat Steps 6 and 7.

Table of Contents