EasyManua.ls Logo

Gilson 157 - Baseline Drift; LAN (Ethernet) Connection

Default Icon
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gilson 157/159 UV-VIS Detectors User’s Guide 5-3
Troubleshooting
5
Troubleshooting
Baseline Drift
Maintain constant temperature conditions during the measurement.
Allow the lamp to warm up for at least 15–30 minutes.
Eliminate changes in ambient temperature. If drift is dependent on temperature, eliminate drafts and avoid
areas of direct sunlight.
Check for contaminated or “bleeding” column. If drift disappears after removing column connections,
regenerate or replace column.
Be sure solvents are degassed.
If you see baseline deflection when back pressure is exerted on the flow cell, you probably have bubbles
trapped in the cell. Try one of these methods to eliminate bubbles:
Increase back pressure on the cell by connecting pressure regulator tubing to the flow cell outlet.
Increase back pressure by raising flow rate and using a syringe or septum to momentarily block exit
stream.
Flush flow cell using a syringe filled with fresh solvent.
Raise the fraction collector to a level higher than the flow cell.
Check pump or delivery system to determine that flow rate is constant.
The lamp intensity may be weak. Refer to Lamp Replacement
on page 4-9.
Check flow cell for contamination. Refer to Flow Cell Maintenance
on page 4-2.
LAN (Ethernet) Connection
For communication to occur, the instrument must be connected via a LAN (Ethernet) connection to a router. If
there are communication problems between the instrument and the computer, verify that the procedure below
has been followed.
1 Connect the router to a power source and then wait 30 seconds for the router to initialize.
2 With the instrument powered OFF, connect one end of an Ethernet cable to a port on the router and the other
end to the ETHERNET port on the rear panel of the instrument.
3 Power the instrument ON and then wait 1 minute for it to initialize. The instrument will be assigned an IP
address via DHCP from the router at this time.
4 Using an Ethernet cable, connect one end of the cable to the connector on the computer and the other end to
a port on the router.
Confirm Communication
After the connections have been verified, use the procedure below to confirm communication.
1 Reboot the computer.
2 Turn off power to the router and the detector.
3 Wait five minutes.
4 Turn power on to the router.
5 Turn power on to the detector.
6 Start TRILUTION® LC Software.

Table of Contents

Related product manuals