GSE Scale Systems - 227
Chapter 21 Setpoints Setup (FEATURE) Technical Reference Manual
<ENTER>
P5112.X Macro ā0ā
<5> <1> <1> <3> <SELECT> <0>
<ENTER>
P5113.0 Mot'n Ign'd
<5> <1> <1> <4> <SELECT> <0>
<ENTER>
P5114.0 Basis new #
<5> <1> <1> <5> <SELECT> <3>
<ENTER>
P5115. AL: 3.
c. Choose proper selections for de-activation of
Setpoint #1:
<5> <1> <3> <0> <SELECT> <1>
<ENTER>
P5130.1 DeAct Below
<5> <1> <3> <1> <SELECT> <0>
<ENTER>
P5131.0 hold 0.0 S
<5> <1> <3> <2> <SELECT> <16>
<ENTER>
P5132.X Macro none
<5> <1> <3> <3> <SELECT> <0>
<ENTER>
P5133.0 Mot'n Ign'd
<5> <1> <3> <4> <SELECT> <0>
<ENTER>
P5134.0 New #:
<5> <1> <3> <5> <SELECT> <2>
<ENTER>
P5135. DL: 2.
<5> <1> <3> <6> <SELECT> <1>
<ENTER>
d. Select the parameter that Setpoint #1 is based
upon:
<5> <1> <5> <0> <SELECT> <30>
<ENTER>
P5150.0 Par 30 QTY
e. Setpoint Setup is now complete. Proceed to
setup macro 0.
21.4 Example Setups
The following setup examples are structured such as an
ASCII file would be. Entering these setups manually
from the front keypad is possible but approached slightly
different. A remote keyboard or a terminal would make
manual entering of the setup a little easier. If the file
were in ASCII form, the setup procedure would be the
quickest of all approaches. GSE recommends all custom
setups should be backed up on a PC and diskette. This
will make service calls and resale of the same program
much easier.
The 574 does not have setpoint outputs, so most
setpoints will invoke macros to perform the necessary
operations required for the application. ie. Setpoints
would normally control valves, lights, buzzers etc.
Macros can prompt the operator with a message on the
display or give an audible tone.
Example #1: Over/Under Indicator
100%s23640%i%e Access Setup Modes,
Allowing Changes
SET LOWER ACTIVATION SETPOINT LEVEL
(ACTIVE BELOW 15 LBS.)
5100%s1%e P5100.1 SPt 1 Outpt
5110%s1%e P5110.1 Activ Below
5111%s0%e P5111.0 hold 0.0 S
5112%s0%e P5112.0 Macro 0
5113%s0%e P5113.0 Mot'n Ign'd
5114%s0%e P5114.0 Basis new #
5115%s15.000000%e P5115. AL: 15.
5130%s0%e P5130.0 DeAct Above
5131%s0%e P5131.0 hold 0.0 S
5132%s16%e P5132.X Macro none
5133%s0%e P5133.0 Mot'n Ign'd
5134%s1%e P5134.1 Basis %% val
5135%s100.000000%e P5135. DL: 100.
5136%s1%e P5136.1 Based SPt 1
5137%s0%e P5137.0 Value AL
5150%s0%e P5150.0 Par 0 Gross
SET UPPER ACTIVATION SETPOINT LEVEL
(ACTIVE ABOVE 25 LBS.)
5200%s1%e P5200.1 SPt 2 Outpt
5210%s0%e P5210.0 Activ Above
5211%s0%e P5211.0 hold 0.0 S
5212%s1%e P5212.1 Macro 1
5213%s0%e P5213.0 Mot'n Ign'd
SECTION - 21.4