GSE Scale Systems - 265
Chapter 28 Information Parameters Technical Reference Manual
This parameter is used to change ALL parameters
including calibration to their factory default values.
Press <ENTER> twice to default the selections.
Changes must be saved when leaving the Setup Mode for
this operation to take effect! Press <SELECT> to see
the next parameter.
Defaulting Parameters and the Model 574 System
When the default setup modes P65000 and P65001 are
invoked, the model number parameter P160 determines
the model number and the subsequent default values. If
the model number is the 574, then the Model 574 default
parameters are used. If the model number is the Model
550, then Model 550 default parameters are used.
P65002.Deflt - CAL
This section works the same as P65001 above except that
all the calibration parameters are retained. Press
<ENTER> twice to default the selections. Changes
must be saved when leaving the Setup Mode for this
operation to take effect! When you are finished
reviewing these informational parameters press
<ZERO> to return directly to the Weigh Mode.
P65010 dbase Reset
Press <ENTER> to simply default the database setups.
Whether or not any database memory is allocated to
general usage is not affected. However if it is desired to
allocate some of the database memory for general use,
press <2> <ENTER>. This will reduce the total amount
of database memory by 8K bytes. To reverse this and
keep all the database module's memory allocated to
databases, press <1> <ENTER>.
Allocating database memory to general usage can be
used to allow variably sized setup modes (such as
custom transmit setup and macro setup) to increase when
they would otherwise be limited by the occurence of the
"Out Of Ram" error message. If this is done, the only
limitation on the amount of information entered into
these setup modes is the amount of available EEPROM
memory.
SECTION - 28.9