CONFIGURATION
107
Rear Panel USB
State
Displays the rear panel USB-B port state. This
setting is not configurable.
F-21
0 = Absent, 2 = USB-CDC,
3 = GPIB-USB adapter
Rear Panel USB
Mode
Sets the rear panel USB mode.
Note: Option #3, USB CDC Full Speed Only,
can be used to reduce the data transmission
speed when there are sources of interference in
the operating environment. This option is only
available for firmware version 1.42 and above.
F-22
0 = Disable, 1 = GPIB-USB
adapter, 2 = Auto detect speed,
3 = Full speed only
GPIB Address
Sets the GPIB address.
F-23
0~30
LAN Settings
MAC Address-
1~6
Displays the MAC address 1~6. This setting is
not configurable.
F-30~F-35
0x00~0xFF
LAN
Turns Ethernet on or off.
F-36
0 = Disable, 1 = Enable
DHCP
Turns DHCP on or off.
F-37
0 = Disable, 1 = Enable
IP Address-1~4
Sets the default IP address. IP address 1~4
splits the IP address into four sections.
(F-39 : F-40 : F-41 : F-42)
(0~255 : 0~255 : 0~255 : 0~255)

Table of Contents

Other manuals for GW Instek PSW Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSW Series and is the answer not in the manual?

GW Instek PSW Series Specifications

General IconGeneral
Ripple & Noise (CC)≤ 3mArms
ProtectionsOVP, OCP, OTP
Remote SenseYes
ProgrammableYes
Operating Temperature0°C to 40°C
Storage Temperature-20°C to 70°C
Voltage Resolution1mV
Current Resolution1mA
Current Accuracy±0.1% + 10mA
Ripple & Noise (CV)≤1mVrms
Load Regulation (CC)≤0.1% + 5mA
Line Regulation (CC)≤0.1% + 5mA
Remote InterfaceUSB, RS232
DisplayLCD