Chapter 5
5. Maintenance and Support
5.1. Routine Maintenance
The Heska Element COAG
Analyzer only requires cleaning on a regular basis as
maintenance. A visual inspection of the instrument for damage and a regular cleaning of
the surface are strongly recommended.
i
Do not use the Heska Element COAG
Analyzer if it appears to be damaged.
Disconnect the power, and contact your Distributor for technical support.
5.2. Cleaning the Outer Shell
Visually inspect the instrument and clean if necessary. Pay special attention to the
cartridge slot opening. Removal of dried blood, dirt or other foreign matters is done by
using a water-moistened cotton swab or a damp cloth. Make sure to remove residual
water with a dry cotton swab or a dry cloth. If a disinfectant is needed, use Virkon (
1 %
solution). After disinfection of the instrument, wipe it with a damp water-moistened cloth
or cotton swab to remove any residual disinfectant.
Use of solvents or strong cleaning agents is inadvisable, as they may damage the
instrument’s surface and the LCD touch screen.
Use of aerosols or pouring detergents, solutions or any other liquid directly onto
the instrument is likewise not recommended.
!
We strongly suggest that you follow the instructions to avoid damaging the
instrument. Damages incurred due to improper maintenance are not covered
by the warranty.
5.3. Cleaning the Cartridge Slot and Air Filter
Cleaning of the cartridge slot and air filter should be done by lightly vacuuming their
openings. The air filter is located on the right side (as seen from the front) of the instrument
and the cartridge slot in front.