3. Configuring the Heska Element COAG
Analyzer
i
A Note on the Communications Log: In the service page of the Heska Element
COAG
Analyzer, a log of the data export communication can be found (the
last 1000 messages are shown).
The log page is intended for LIS service technicians to find transmission prob-
lems when integrating the Heska Element COAG
Analyzer into their LIS
system.
The transmitted data is shown at protocol layer with a primary column indicating
direction to and from the instrument.
In addition to the transmitted data, the Heska Element COAG
Analyzer will
log messages when the communication with the server is established or lost.
Also, miscellaneous error messages will appear here.
3.2.7. Quality Assurance
Figure 3.16:
Quality
assurance menu.
3.2.7.1. Calibration
The Calibration menu is shown in
Figure 3.18
. The Calibration menu gives an overview
over date(s) and time(s) for when calibration for the Fibrinogen Test has been performed.
The currently active calibration is indicated with a dark purple button to the right of the
row with the calibration information.