i
A Note on the Communications Log: In the service page of the Heska Element
COAG
™
Analyzer, a log of the data export communication can be found (the
last 1000 messages are shown).
The log page is intended for LIS service technicians to find transmission prob-
lems when integrating the Heska Element COAG
™
Analyzer into their LIS
system.
The transmitted data is shown at protocol layer with a primary column indicating
direction to and from the instrument.
In addition to the transmitted data, the Heska Element COAG
™
Analyzer will
log messages when the communication with the server is established or lost.
Also, miscellaneous error messages will appear here.
3.2.7. Quality Assurance
Figure 3.16:
Quality
assurance menu.
3.2.7.1. Calibration
The Calibration menu is shown in
Figure 3.18
. The Calibration menu gives an overview
over date(s) and time(s) for when calibration for the Fibrinogen Test has been performed.
The currently active calibration is indicated with a dark purple button to the right of the
row with the calibration information.