8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:CSET:PARAmeter data1,data2
dat
a
When the comparison method of the comparator is set as high limit and low
limit values in resistance measurement mode:
data1: high limit value, data2: low limit value
Both data1 and data2 = 0 to 99999 (numerical data in NRf format)
When the comparison method of the comparator is set as high limit and
low limit values in temperature conversion mode:
Both data1 and data2 = ± 999.9 (numerical data in NRf format)
When the comparison method of the comparator is set as standard value
and range:
data1 : standard value, data2 : range
data1 = 0 to 99999 (numerical data in NRf format)
data2 = 0.00 to 99.99 (numerical data in NRf format)
Functio
n
Sets the comparator range of the comparator set by ":CSET:TABLe"
command.
Erro
r
When low limit value bigger than high limit value is set, the execution
error is generated.
Exampl
e
Transmission
Transmission
:RESI:RANG 300;:CSET:CMOD HL;PARA 10100,9900
:RESI:RANG 300;:CSET:CMOD REF;PARA 10000,1.0
These examples executes the comparator in the range of 99.00Ω to
101.00Ω.
Synta
x
:CSET:PARAmeter?
Functio
n
Return the comparator range of the comparator set by ":CSET:TABLe"
command, as a numerical value in NRf format (data).
Respons
e
If headers are ON
If headers are OFF
:CSET:PARAMETER data1,data2
data1,data2
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:CSET:PARA?
:CSET:PARAMETER 10100,9900
:CSET:PARA?
10100,9900
:CSET:PARAmeter
:CSET:PARAmeter?