93
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:DELTa:MODE data
dat
a
DELTa, TEMPerature (character data)
Functio
n
Sets the display method of the temperature conversion measurement.
Erro
r
The execution error is generated when not in the temperature conversion
display.
Exampl
e
Transmission
:DELT:MODE DELT
The temperature conversion display is set to the Δt display.
Synta
x
:DELTa:MODE?
dat
a
Returns the current display method of the temperature conversion setting,
as DELTA or TEMPERATURE data.
Respons
e
If headers are ON
If headers are OFF
:DELTA:MODE data
data
Erro
r
When not in the temperature conversion display, the execution error is
generated.
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:DELT:MODE?
:DELTA:MODE DELTA
:DELT:MODE?
DELTA
:DELTa:MODE
:DELTa:MODE?