94
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
DELTa:TABLe data
dat
a
0 to 15 (numerical data in NR1 format)
Functio
n
In the temperature conversion mode, specifies the temperature conversion
table as followa;
When data is 0, sets the temperature conversion display to be off.
When data is 1 to 15, uses the designated table number for the temperatur
e
conversion display.
Erro
r
The execution error is generated in the following circumstances.
When the temperature correction function is selected at power on.
When the measurement mode is not set to the resistant measurement
mode.
When the designated table number is invalid.
When one of the values 1 to 15 is specified and no TC sensor is connected.
Exampl
e
Transmission
DELT:TABL 1
The temperature conversion table number 1 is used.
Synta
x
:DELTa:TABLe?
Functio
n
Returns current temperature conversion table, as 0 to 15 (data).
The setting conditions are as follows.
When data is 0, sets the temperature conversion display to be OFF.
When data is 1 to 15, shows the temperature conversion table number that
is used.
Respons
e
If headers are ON
If headers are OFF
:DELTA:TABLE data
data
Erro
r
The executed error is generated in the following circumstances.
When the temperature correction function is selected at power on.
The measurement mode is not set to the resistance measurement mode.
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:DELT:TABL?
:DELT:TABL 1
:DELT:TABL?
1
:DELTa:TABLe
:DELTa:TABLe?