95
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:DSET:AUTO
Functio
n
Writes the current measurement values (resistance and temperature) and
coefficient (235.0) automatically into the temperature conversion table
specified with the "DSET:TABLe" command.
Erro
r
When the temperature correction function is selected at power on, the
executed error is generated.
Exampl
e
Transmission
:DSET:TABL 1;AUTO
Writes the current measurement values and the coefficient into the
temperature conversion table.
Synta
x
:DSET:PARAmeter data1, data2, data3
dat
a
data1: initial resistance, data2: initial temperature, data3: constant
data1 = 0 to 999.99E3 (numerical value in NRf format)
data2 = -10.0 to 99.9 (numerical value in NRf format)
data3 = ±999.9 (numerical value in NRf format)
Functio
n
Sets the initial resistance, initial temperature, and constant of the
temperature conversion table designated by ":DSET:TABLe" command.
Erro
r
When the temperature correction function is selected at power on, the
executed error is generated.
Exampl
e
Transmission
:DSET:TABL 1;PARA 300.00E‑3,20.0,235.0
In the first temperature conversion table, this sets the initial resistance
value to 300 mΩ, the initial temperature to 20, and the coefficient to the
constant for upper.
:DSET:AUTO
:DSET:PARAmeter