103
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:RESIstance:AUTO?
Functio
n
Returns the current auto range setting, as ON or OFF (data).
Respons
e
If headers are ON
If headers are OFF
:RESISTANCE:AUTO data
data
Erro
r
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:RESI:AUTO?
:RESISTANCE:AUTO ON
:RESI:AUTO?
ON
Synta
x
:RESIstance:RANGe data
dat
a
0.3, 3, 30, 300, 3000, 30000, 300000 (numerical value in NRf format)
Functio
n
Sets the resistance measurement range.
Not
e
For this command, the measurement range is set to "the designated value
measurable range". However, when the command exceeds the full scale of
the range, it is set to an upper range.
Erro
r
The execution error is generated in the following circumstances.
When the value over the maximum range or the minus value is designated
.
When the measurement mode is set to the temperature measurement.
When designated 30 kΩ and 300 kΩ range in case that the measurement
sampling rate is set to FAST.
When in the temperature conversion display
Exampl
e
Transmission
:RESI:RANG 300.005
The resistance measurement range is set to the 3 kΩ range.
:RESIstance:AUTO?
:RESIstance:RANGe

Table of Contents