103
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
Returns the current auto range setting, as ON or OFF (data).
Respons
If headers are ON
If headers are OFF
:RESISTANCE:AUTO data
data
Erro
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
Transmission
Response
:RESI:AUTO?
:RESISTANCE:AUTO ON
:RESI:AUTO?
ON
Synta
:RESIstance:RANGe data
dat
0.3, 3, 30, 300, 3000, 30000, 300000 (numerical value in NRf format)
Functio
Sets the resistance measurement range.
Not
For this command, the measurement range is set to "the designated value
measurable range". However, when the command exceeds the full scale of
the range, it is set to an upper range.
Erro
The execution error is generated in the following circumstances.
・ When the value over the maximum range or the minus value is designated
・ When the measurement mode is set to the temperature measurement.
・ When designated 30 kΩ and 300 kΩ range in case that the measurement
sampling rate is set to FAST.
・ When in the temperature conversion display
Exampl
Transmission
:RESI:RANG 300.005
The resistance measurement range is set to the 3 kΩ range.
:RESIstance:AUTO?
:RESIstance:RANGe