104
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
:RESIstance:RANGe?
Functio
Returns the current resistance measurement range setting, as a numerical
value in NRf format (data).
Respons
If headers are ON
If headers are OFF
:RESISTANCE:RANGE data
data
Erro
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
Transmission
Response
:RESI:RANG?
:RESISTANCE:RANGE 300E0
:RESI:RANG?
300E0
Synta
"FAST, MEDIum, SLOW (character data)
Functio
Sets the measurement sampling rate.
Not
When the sampling rate is set to FAST, in case that the resistance
measuring range is set to 30 kΩ or 300 kΩ range, the resistance
measuring range is set to 3 kΩ range automatically.
Erro
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
Transmission
:SAMP MEDI
The measurement sampling rate is set to MEDIUM.
:RESIstance:RANGe?
:SAMPle