104
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:RESIstance:RANGe?
Functio
n
Returns the current resistance measurement range setting, as a numerical
value in NRf format (data).
Respons
e
If headers are ON
If headers are OFF
:RESISTANCE:RANGE data
data
Erro
r
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:RESI:RANG?
:RESISTANCE:RANGE 300E0
:RESI:RANG?
300E0
Synta
x
:SAMPle data
dat
a
"FAST, MEDIum, SLOW (character data)
Functio
n
Sets the measurement sampling rate.
Not
e
When the sampling rate is set to FAST, in case that the resistance
measuring range is set to 30 kΩ or 300 kΩ range, the resistance
measuring range is set to 3 kΩ range automatically.
Erro
r
When the measurement mode is set to the temperature measurement, the
execution error is generated.
Exampl
e
Transmission
:SAMP MEDI
The measurement sampling rate is set to MEDIUM.
:RESIstance:RANGe?
:SAMPle