81
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
*
ESR?
Functio
n
The contents of the standard event status register SESR are returned as an
numerical value (data) in NR1 format in the range 0 to 255, and clears its
contents.
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
PON URQ CME EXE DDE QYE RQC OPC
Respons
e
Whether headers are on or off
data
Exampl
e
Transmission
Response
*ESR?
32
Synta
x
*
IDN?
Functio
n
The response consists of the name of the manufacturer of the unit, the
model name, and the software version.
Not
e
The
*
IDN? query is the last query message of the program messages.
Accordingly, if another query is detected after this query, a query error is
generated, and no response message after the
*
IDN? query is produced.
Respons
e
Whether headers are on or off
HIOKI,3227,0,V2.00
First field: Manufacturer’s name
Second field: Model name
Third field: Not used - always zero
Fourth field: Software version
Exampl
e
Transmission
Response
*IDN?
HIOKI,3227,0,V2.00
*
ESR?
*
IDN?