8.6 Message Reference
132
Reference Resistance Setting
Syntax
Command
Query
Response
:CALCulate:LIMit:REFerence <Reference Resistance>
:CALCulate:LIMit:REFerence?
<Reference Resistance>
<Reference Resistance> = 0 to 999999 (NR1)
Example
Command
:CALC:LIM:REF 141000
Note
Reference Resistance is specified as an integer value.2. To specify 0.567 Ω in the
2Ω range, send the following command:
:CALCulate:LIMit:REFerence 56700
Decision Tolerance Setting
Syntax
Command
Query
Response
:CALCulate:LIMit:PERCent <Tolerance (%)>
:CALCulate:LIMit:PERCent?
<Tolerance (%)>
<Tolerance (%)> = 0 to 99.999 (NR2)
Example
Command
:CALC:LIM:PERC 10.000
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
:CSET:PARAmeter
Comparator Result
Syntax
Query
Response
:CALCulate:LIMit:RESult?
<HI, IN, LO, OFF or ERR>
Example
Query
Response
:CALC:LIM:RES?
HI
Setting and Querying BIN Measurements
BIN Measurement State
Syntax
Command
Query
Response
:CALCulate:BIN:STATe <1, 0, ON or OFF>
:CALCulate:BIN:STATe?
<ON or OFF>
Example
Command
:CALC:BIN:STAT ON