8.6 Message Reference
134
(5) System
Reference Resistance Setting
Syntax
Command
Query
Response
:CALCulate:BIN:REFerence
<BIN No.>,<Reference Resistance>
:CALCulate:BIN:REFerence? <BIN No.>
<Reference Resistance>
<BIN No.> = 0 to 9
<Reference Resistance> = 0 to 999999 (NR1)
Note
Reference Resistance is specified as an integer value.
To specify 0.567 Ω in the 2Ω range, send the command as follows:
:CALCulate:BIN:REFerence 5,56700
Decision Tolerance Setting
Syntax
Command
Query
Response
:CALCulate:BIN:PERCent <BIN No.>,<Tolerance (%)>
:CALCulate:BIN:PERCent? <BIN No.>
<Tolerance (%)>
<BIN No.> = 0 to 9
<Tolerance (%)> = 0 to 99.999 (NR2)
Query the Decision Result
Syntax
Query
Response
:CALCulate:BIN:RESult?
<NR1>
<NR1> = 0 to 1023
The bit corresponding to each BIN with a PASS decision is set to 1.
Example
Query
Response
:CALC:BIN:RES?
128
BIN7 was judged PASS.
Offset Voltage Compensation State
Syntax
Command
Query
Response
:SYSTem:OVC <1, 0, ON or OFF>
:SYSTem:OVC?
<ON or OFF>
Example
Command
:SYST:OVC ON
Query
Response
:SYST:OVC?
OFF
Note
Settings in the 110 KΩ range and higher are ignored.
bit 9 bit 8 bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
BIN9 BIN8 BIN7 BIN6 BIN5 BIN4 BIN3 BIN2 BIN1 BIN0