135
8.6 Message Reference
Temperature Measurement Settings (Analog Input)
Temperature Sensor Selection
Syntax
Command
Query
Response
:SYSTem:TEMPerature:SENSor <PT or ANALog>
:SYSTem:TEMPerature:SENSor?
<PT or ANALOG>
PT ...............The 9451 TEMPERATURE PROBE is used as the
temperature sensor
ANALOG ....An analog output thermometer is used as the
temperature sensor
Example
Command
:SYST:TEMP:SENS ANAL
Query
Response
:SYST:TEMP:SENS?
PT
Note For some commands, RS-232C cannot be selected as input for
temperature measurement. In such a case, use the menu screens on the
3541 unit to make the setting.
❖ Temperature measurement via RS-232C interface (using the 3444/3445
TEMPERATURE HiTESTER+ 3909 INTERFACE PACK) (Page 34)
Analog Input Parameter Settings
Syntax
Command
Query
Response
:SYSTem:TEMPerature:PARameter <V1>,<T1>,<V2>,<T2>
:SYSTem:TEMPerature:PARameter?
<V1>,<T1>,<V2>,<T2>
<V1> = 0 to 2.00 (NR2)............. Reference Voltage 1 [V]
<T1> = -99.9 to 999.9 (NR2)..... Reference Temperature 1 [°C]
<V2> = 0 to 2.00 (NR2)............. Reference Voltage 2 [V]
<T1> = -99.9 to 999.9 (NR2)..... Reference Temperature 2 [°C]
Example
Command
:SYST:TEMP:PAR 0,-10,2,100
Query
Response
:SYST:TEMP:PAR?
0.00,0.00,1.00,100.0
0 V displays as 0°C, and 1 V displays as 100°C.