8.6 Message Reference
136
Self-Calibration State and Setting
Execute Self-Calibration
Syntax
Command
:SYSTem:CALibration
Set Self-Calibration Execution State
Command
Query
Response
:SYSTem:CALibration:AUTO <1, 0, ON or OFF>
:SYSTem:CALibration:AUTO?
<ON or OFF>
ON......... AUTO Self-Calibration selected
OFF ...... MANUAL Self-Calibration selected
Example
Command
:SYST:CAL:AUTO OFF
Query
Response
:SYST:CAL:AUTO?
ON
Note
Even when AUTO is selected, Self-Calibration can be manually performed at any
time by sending the SYSTem:CALibration command.
Set and Query the Key Beeper Setting
Syntax
Command
Query
Response
:SYSTem:BEEPer:STATe <1, 0, ON or OFF>
:SYSTem:BEEPer:STATe?
<ON or OFF>
Example
Command
Query
Response
:SYST:BEEP:STAT ON
:SYST:BEEP:STAT?
ON
Select and Query the Line Frequency Setting
Syntax
Command
Query
Response
:SYSTem:LFRequency <50 or 60>
:SYSTem:LFRequency?
<50 or 60>
Example
Command
Query
Response
:SYST:LFR 50
:SYST:LFR?
60
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
:FREQuency