137
8.6 Message Reference
Set and Query the Key-Lock State
Syntax
Command
Query
Response
:SYSTem:KLOCk <1, 0, ON or OFF>
:SYSTem:KLOCk?
<ON or OFF>
Example
Command
Query
Response
:SYST:KLOC ON
:SYST:KLOC?
OFF
Save and Load Measurement Setting States
Syntax
Command
:SYSTem:SAVE <Table No. 1 to 30>
:SYSTem:LOAD <Table No. 1 to 30>
Set and Query the Header Present Setting
Syntax
Command
Query
Response
:SYSTem:HEADer <1, 0, ON or OFF>
:SYSTem:HEADer?
<ON or OFF>
Example
Command
Query
Response
:SYST:HEAD ON
:SYST:HEAD?
OFF
:SYSTEM:HEADER ON
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
HEADer
Select the ERR Output Setting
Syntax
Command
Query
Response
:SYSTem:ERRor <SYNChronous or ASYNchronous>
:SYSTem:ERRor?
<SYNChronous or ASYNchronous>
SYNCHRONOUS ......Synchronize with EOC output
ASYNCHRONOUS ....Asynchronous with EOC output
Example
Command
:SYST:ERR SYNC
Query
Response
:SYST:ERR?
ASYNCHRONOUS