EasyManua.ls Logo

Hioki 3541 - Page 144

Hioki 3541
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.6 Message Reference
138
BCD Output Setting
Syntax
Command
Query
Response
:SYSTem:EXTernalout <BIN or BCD>
:SYSTem:EXTernalout?
<BIN or BCD>
Example
Command
:SYST:EXT BCD
Query
Response
:SYST:EXT?
BIN
Note
BIN output is disabled when BCD output is selected.
BCD output is disabled when BIN output is selected.
Delimiter Setting
Syntax
Command
Query
Response
:SYSTem:TERMinator <0 or 1>
:SYSTem:TERMinator?
<0 or 1>
0 ....... LF+EOI
1 ....... CR,LF+EOI
Example
Command
:SYST:TERM 1
Query
Response
:SYST:TERM?
0
Note At power-on, this is set to 0 (LF+EOI).
The RS-232C delimiter is fixed as CR + LF.
System Reset
Syntax
Command
:SYSTem:RESet
Description
Command Returns all settings, including any saved data, to factory default
settings.
Example
Command
:SYST:RES
Note
If you want to preserve saved data, use the RST command instead.

Table of Contents