5.2 BIN Measurement Function
52
6
Switches to comparison method selection for measurements.
Select the comparison method. Each key-press changes the displayed
selection.
7
Switches the display to upper/lower threshold setting
Set the upper and lower thresholds.
8
Returns to the Main Display of this BIN No.
Repeat Steps 3 to 7 for each BIN No.
9
Applies setting and returns to the Measurement display.
The BIN function is enabled.
To cancel the settings:
Compare with
reference value/
tolerance
Compare with
upper/low
thresholds
(Sub Display)
In this case, select HIGH/LOW.
Or ten-keys
(Sub Display)
(Main Display)
In this case, set the
upper threshold to
1000 and the lower
threshold to
800 .
Upper threshold
Lower threshold
Select a digit
Select numerical value
Select a digit to change by moving
the blinking location, then select the
new numerical value.
Using the RANGE keys:
Using the ten-keys:
Press the numeric keys
corresponding to the digits to
be entered.
In this case, select BIN2 as
shown at the left.
BIN2
Upper threshold: 900
Lower threshold: 700