63
5.6 Statistical Calculation Functions
Enabling/Disabling the Statistical Calculation Function, and Clearing
Calculation Results
1
The Statistical Calculation display appears.
2
The function enable/disable display appears.
Enable or disable the calculation function on the Sub Display.
on ........ enables the calculation function (ON).
oFF ...... disables the calculation function (OFF).
CLr....... clears calculation results.
3
Applies setting and returns to the Measurement display.
To cancel the settings:
(Sub Display)
(Main Display)
(Sub Display)
• Statistical Calculation function setting (ON, OFF, CLR) is not
available when the Comparator or BIN function is enabled.
• If Statistical Calculation is turned off and then back on without first
clearing calculation results, it resumes calculating from the point
when it was turned off.
• The Statistical Calculation function slows measurements when it is
ON.
Importing Data
Pressing the TRIG key while Statistical Calculation is ON executes
one of the following operations:
• External Trigger: Takes one measurement and performs statistical
calculation on the result
• Internal Trigger: Performs statistical calculation on the value
displayed immediately after pressing
• ∗TRG command executes the same operation.
• Grounding the TRIG
terminal of the EXT I/O connector executes