Usage Notes
10
Before Connecting Measurement Cables
To avoid electric shock, turn off the power to all devices before plugging or
unplugging any cables or peripherals.
Be sure to connect the voltage input and current input terminals correctly. An
inc
orrect connection could damage or short circuit this instrument.
In order to prevent electric shock or device damage, observe the following pre-
cautions when making connections to the external control terminals and other
int
erface connectors.
•Turn off the instrument and any equipment being connected before connect-
ing the measurement cables.
•Exercise care not to exceed the ratings of external control terminal and inter-
face connector signals.
•Make connections securely to avoid the risk of connections coming loose
during ins
trument operation and bringing wires into contact with other electri-
cally conductive parts.
•Ensure that devices and systems to be connected to the external control ter-
minals are properly isolated.
To avoid electric shock and short-circuit accidents, use only the supplied voltage
cords to connect the instrument input terminals to the circuit to be tested.
To avoid equipment failure, do not disconne
ct the communications cable while com-
munications are in progress.
Use a common ground for both the instrument and the computer. Using different
ground
circuits will result in a potential difference between the instrument's ground
and the computer's ground. If the communications cable is connected while such a
potential difference exists, it may result in equipment malfunction or failure.
Before connecting or disconnecting any communications cable, always turn off the
in
strument and the computer. Failure to do so could result in equipment malfunction
or damage.
After connecting the communications cable
, tighten the screws on the connector
securely. Failure to secure the connector could result in equipment malfunction or
damage.
HIOKI PQ3198A961-02