13.3 Screen Specifications
216
(2) Hardware settings
Display language Japanese/English/Chinese Simple (Simplified)/Chinese Trad (Traditional)/Korean/German/
French/Italian/Spanish/Turkish/Polish
Beep sound ON/OFF
Screen color COLOR1/COLOR2/COLOR3/COLOR4/COLOR5
Clock setting Western calendar year, month, day, hours, and minutes
LCD backlight AUTO OFF (2 min) /ON (Continuous)
Backlight automatically turns off 2 min. after last key operation.
Once the backlight has turned off, it will automatically turn back on with operatio
n of any key (in-
cluding when the key lock is engaged).
System reset System reset reverts the instrument to factory defaults (except for display language, time, phase
names, RS host, IP address, subnet mask, default gateway and FTP server settings).
Instrument information Software version and serial number display
External event output parame-
ters
OFF/short pulse/long pulse/ ΔV10 alarm (when ΔV10 has been selected during flicker measure-
ment)
External control (IN) Event, START/STOP
ΔV10 alarm threshold 0.00 V to 9.99 V
External interface settings
RS-232C RS host: OFF/GPS
GPS: Time zone, expressed as the divergence from coordinated universal time
(UTC): -13:00 to +13:00, user-selectable
LAN
IP Address : 3 characters.3 characters.3 characters.3 characters (***.***.***.***)
Subnet Mask : 3 characters.3 characters.3 characters.3 characters (***.***.***.***)
Default Gateway: 3 characters.3 characters.3 characters.3 characters (***.***.***.***)
FTP authentication: ON/OFF
User name : Up to 20 one-byte characters (Available only with the authentication set
to on)
Password : Up to 20 one-byte characters (Available only with the authentication set
to on)
(3) Recording Settings
Time Start Manual/Time/Exactly
Start time and date: Western year/Month/Day Hours:Minutes
Stop time and date: Western year/Month/Day Hours:Minutes
(The stop time cannot be set when repeat recording is set to 1 week. If repeat recording is set to
1 day, the hours and minutes can be set based on the start and stop times.)
Repeat setting OFF/1 Week/1 Day
OFF: Repeat recording not performed.
1 Week: Repeat recording is performed one week at a time. Set the repeat count.
1 Day: Repeat recording is performed one day at a time. Specify the start and stop times for one
day.
Repetition time With a repeat setting of one day, specify the start and stop times for one day.
Start time: Hours and minutes, in 1-minute increments (using 24-hour time)
Stop time: Hours and minutes, in 1-minute increments (using 24-hour time)
Repeat number When repeated recording is set to 1 week: 1 to 55 count
When repeated recording is set to 1 day: 1 to 366 count (When
actual time control is enabled, set
based on the stop time and date.)
HIOKI PQ3198A961-02