5.3 Changing the Measurement Period
80
Recording starts when the START/STOP key is
Relationship between the repeat setting and the maximum repeat count
When the repeat setting is [OFF]
Recording stops when the START/STOP key is pressed, or at
the stop time and date set for real-time control.
Max. 35 days
Data storage folder
Measurement data and event data
Stand
by
When the repeat setting is [1 Day]
The START/STOP key is pressed.
No gap between the start time and stop time settings, e.g., if the start time
and stop time settings are both set to 0:00 (or any same time).
Stan
dby
1 day 1 day 1 day 1 day 1 day
Recording
stops.
Recording
stops.
Recording
starts.
Recording
starts.
Max. 366
days
Measurement data and event data
Data storage folder
When the repeat setting is [1 Week]
Recording starts when the START/STOP key is pressed.
No gap No gap No gap No gap No gap
1 week 1 week 1 week 1 week 1 week
Measurement data and event data
Data storage folder
Recording stops.
Recording starts.
Start time /
Stop time
Max. 55
Weeks
Recording
starts.
Recording stops.
Recording
starts.
Recording
stops.
n 366 repeat count
n 55 repeat count
pressed.
For more information about the data storage folder hierarchy, see "File structure
(overall)" (p.162).
In the event of a power outage (interruption
of power to the instrument), the folder will
be segmented.
Once the data storage files exceed about 1
00 MB, data will be segmented, regard-
less of the repeat count.