10.7 Remote Control of the Instrument by Internet Browser
163
Chapter 10 Using Communications (USB/LAN)
10
The instrument includes a standard HTTP server function that supports remote con-
trol by an internet browser on a computer.
The instrument's display screen and control panel keys are emulated in the browser.
Operating procedures are the same as on the instrument.
10.7 Remote Control of the Instrument by Inter-
net Browser
It is recommended to use either Microsoft Internet Explorer ver-
sion 8 or later.
Set the browser security level to “Medium” or “Medium-high,” or
enable
Active Scripting settings.
Unintended operations may occur if remote control is attempted
from multiple
computers simultaneously. Use one computer at a
time for remote control.
Remote control can be performed even if the instrument's key
lock is active.
1
Launch Internet Explorer.
2
In the address bar, enter “http://” followed by the IP address with which
you configured the instrument.
For example, enter the address as shown below if the instrument’s IP address is
[192.168.1.31]:
3
If the main page is displayed as shown below, you have successfully
connected to the instrument: