4.3 Changing Recording (Save) Settings
74
Recording start method
Sets the method used to start recording.
Recording 2 Setting Screen
Selection
MANUAL
Starts recording from the point at which the [START/
STOP] key is pressed.
TIME
Recording is started at the set time (YY/MM/DD hh:mm).
If the set time has already passed, the “interval time” start-
ing method is used.
INTERVAL
Starts recording at an even division after the interval time
elapses.
Example:
If the [START/STOP] key is pressed at 10:41:22 with the
save interval set to 30 minutes, the instrument will enter
the standby state, and recording will start at 11:00:00.
Similarly, if the save interval is set to 10 minutes, record-
ing will start at 10:50:00. If the save interval is set to 30
seconds or less, recording will start from the next :00 sec-
onds.
REPEAT
Repeats recording while segmenting files every day.
Recording is started on the repeat start date. If the start
date has already passed, recording is started at the
recording time period start time on the next day.
Recording is stopped when the recording time period on
the stop date ends.