4.8 Viewing Motor Measurement Values (Model PW3390-03 only)
94
Motor analysis can be performed using the PW3390-03 (model with motor analysis and D/A output).
When the motor analysis function is installed, the [Motor] page appears on the Measurement and Setting
screens.
The motor analysis function acquires signals from rotation-sensing devices such as a torque sensor and
rotary encoder and measures motor analysis items (torque, rotation rate, motor power and slip).
Motor efficiency, total efficiency and loss can be calculated when combined with the "4.7 Viewing Effi-
ciency and Loss Measurement Values" (p. 89) functions.
See Motor measurements can be displayed simultaneously with motor input voltage and current, power measure-
ments and motor efficiency. "Selecting Measured Items for Display" (p. 46)
4.8 Viewing Motor Measurement Values
(Model PW3390-03 only)
Setting the Motor Sync Source
Tor q u e
Rotation Rate
Motor Power
Slip
Press and then to select the [Motor] page.
Motor Sync Source Setting CH A Range Setting CH B Range Setting
• When [CH A] measurement units are set to [V] and [Hz], or [CH B] units are set to other than
[r/min], motor power [Pm] display is always disabled (“OFF”).
• When [CH B] measurement units are set to [V], slip cannot be calculated, and
[- - - - - -] is