10.1 General Specifications
189
10
10
Chapter 10 Specifications
Specifications Chapter 10
10.1 General Specifications
Operating environment Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature and humidity -10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation)
Dustproof and waterproof IP30 (EN60529)
With CF card cover open: IP20
Applicable standards Safety: EN61010
EMC: EN61326 Class A
Power supply Rated supply voltage: 100 V to 240 V AC (Voltage fluctuations of ±10% for the rated supply voltage
are taken into account.)
Rated supply frequency: 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
Maximum rated power: 140 VA
Backup battery life Clock, settings and integration values for backup (Lithium battery), Approx. 10 years (at 23°C, as a
referential)
Interface USB (function), USB memory, LAN, CF card, RS-232C, Synchronous control
Dimensions Approx. 340 W ×170 H ×156 D mm (13.39" W ×6.69" H ×6.14" D) (excluding protrusions)
Mass Approx. 4.6 kg (162.3 oz.) when PW3390-03
Product warranty period 3 years
Accessories See "Confirming Package Contents"( p.2)
Options See "Options"( p.3)