8.4 Connecting the Instrument to a LR8410 Link-compatible Logger
172
The instrument can be connected to Hioki model LR8410 Link-compatible logger (LR8410 Wireless Log-
ging Station or LR8416 Wireless Heat Flow Logger) via Bluetooth
, enabling it to send measured values
for D/A output parameters wirelessly to the logger (D/A9 to D/A16, up to 8 parameters). Establishing such
a connection will enable the LR8410 Link-compatible logger to observe and record measured values from
the instrument along with measurement targets such as voltage, temperature, and humidity across multi-
ple channels.
To connect the instrument to an LR8410-compatible logger, you will need the following Bluetooth
serial
conversion adapter and power adapter:
Bluetooth
serial conversion adapter: Parani-SD1000 (from SENA Technologies Co., Ltd.)
Bluetooth
Class 1
AC/DC power adapter: OPA-G01 (from SENA Technologies Co., Ltd.)
Operating precautions
To ensure safety, be sure to turn off the instrument before connecting it to the adapter. Turn on the
instrument after connecting the adapter.
Refer to the Parani-SD1000 operating precautions for more information about Bluetooth

use.
Because values are displayed at the resolution of the logger being used, they will differ slightly from the
measured values that are displayed by the instrument. To record values that are closer to the instru-
ment’s measured values, choose a range that is appropriate for the input.