7.4 Saving Operations
138
7.4 Saving Operations
Save destination CF card
USB flash drive
Saving method
Press .
PW3390
(or any other folder)
M3390000.CSV
M3390001.CSV
An error will be displayed if the sequential file number in the folder
reaches 1000. Set a new save location.
Up to 1,000 files
Manual Saving (p. 139)
M3390000.CSV
• When the storage media is full, saving stops.
Storage me-
dia root folder
Files are named M3390000 to
M3390999.
A new file is cre-
ated when any
of the following
are changed:
Save
destination
folder
Wiring mode
Measurement
objects to be
saved
Save destination CF card
Saving method
Automatically,
according to timing
control settings
PW3390
(or any other folder)
11190000.CSV
11190100.CSV
Up to 100 files per day (the file name
is changed each day, so continuous
saving is supported)
Auto-saving (p. 141)
11190000.CSV
• When an auto-saved file exceeds 100 MB,
a new file is created and saving continues.
• Up to 100 files per day can also be saved to the root folder.
• When the storage media is full, saving stops.
Files are named MMDD0000 to
MMDD9900.
CF card root
This example describes data that was automatically saved on November 19.
Save destination CF card
USB flash drive
Saving method
Press on the
[Wave + Noise]
page of the Measure-
ment screen
PW3390
(or any other folder)
W3390000.CSV
W3390001.CSV
Up to 1,000 files
Waveform Data Saving (p. 145)
W3390000.CSV
• When the storage media is full, saving stops.
Files are named W3390000 to
W3390999.
Storage me-
dia root folder
An error will be displayed if the sequential file number in the folder reaches 1000.
Set a new save location.
Save destination CF card
USB flash drive
Saving method
Display the screen to
save, and press
and .
PW3390
(or any other folder)
H3390000.BMP
H3390001.BMP
Up to 1,000 files
Screen Capture Image Saving (p. 147)
H3390000.BMP
• When the storage media is full, saving stops.
Files are named H3390000 to
H3390999.
Storage me-
dia root folder
An error will be displayed if the sequential file number in the folder reaches 1000.
Set a new save location.
Save destination CF card
USB flash drive
Saving method
Move to a folder on
the File Operations
screen, and press
.
PW3390
(or any other folder)
SETTING1.SET
SETTING2.SET
until the storage media is full
Saving Setting Configurations (p. 149)
• Files can also be saved to the root folder.
• When the storage media is full, saving stops.
The user-specified file name may
be up to 8 characters.
Storage me-
dia root folder
• The maximum number of files to save in the root for manual save, waveform and screen copy
is up to 512 files in FAT16 format.
• Changing the save destination or cycling the instrument resets the sequential numbers. Once
the sequential numbers have been reset, newly created files are numbered sequentially skip-
ping the existing numbers.