5.7 Trend Function
119
5
Chapter 5 Operating Functions
You can select up to eight basic measurement parameters and display fluctuations in their respective
measured values as a graph. The generated screens can be saved as screenshots.
Press and key to display the [Trend] page.
Graphed parameters
The graph number, graphed parameter, and scale setting are shown.
[SEMI AUTO] is shown if a scale factor has been set, while [AUTO] or
[MANUAL] is shown if using the AUTO or MANUAL setting, respectively.
5.7 Trend Function
Displaying the Trend screen
Minimum values on graph
Graphed parameters
Maximum values on graph
Time axis
Waveforms
being graphed
Graph dis-
play method
The displayed graph waveforms are generated by graphing virtual D/A output wave-
forms on the screen. Consequently, some combinations of display parameters may
result in unusually shaped graphs based on the D/A output rules.
• D/A output rules
See: "8.3.3 Output Level" (p.169), "8.3.4 D/A Output Examples" (p.170)
• Integration scale and frequency full scale settings
See: "8.3.2 Output Item Selection" (p.166)
On the Trend screen, measured values that fall outside the display range (see "10.5 Mea-
surement Item Details" (p.207)) are clipped in principle to the maximum display value.
This functionality applies to the following displays:
• Graph waveforms
• Icons that indicate present measured values
To save fluctuations in measured values, use the auto-saving function (p.141) together.