8.3 Using Analog and Waveform D/A Output
171
8
Chapter 8 Connecting External Devices
(1) Analog output changes when integration starts, and is held constant after integration stops.
(2) When the integration value exceeds
±5 V, analog output becomes 0 V and changes proceed from there.
(3) When the data hold is activated during integration, analog output is held constant. However, when
data hold is canceled, analog output returns to the actual integration value.
(4) The integration value is reset, and analog output becomes 0 V.
Output retaining
Stop
Integration
Integration
Reset
Data Hold
Start Integration Resume Integration Resume Integration
Data Hold
Cancel
Start Integration
Updated at each interval
Output retaining
Data Hold
Time
Output retaining
Stop Integration
Stop Integration