Up/Down Controller User Info/Support Blood Glucose Records
Living with Diabetes
9
105
Possible causes of hyperglycemia (high blood glucose)
Possible cause of high Suggested action
Expired insulin Deactivate and remove used pod.
Apply new pod filled with insulin
from a new vial.
Insulin exposed
to extreme
temperatures
Infusion site in or
near a scar or mole
Deactivate and remove used pod.
Apply a new pod in a different
location.
Infected infusion site Deactivate and remove used pod.
Apply a new pod in a different loca-
tion and consult your healthcare
provider.
Dislodged cannula Deactivate and remove used pod.
Apply a new pod in a different
location.
Avoid sites near a waistband,
belt, or other areas where fric-
tion may dislodge the cannula.
Possible cause of high Suggested action
Empty pod Deactivate and remove used pod.
Apply a new pod in a different
location.
Incorrect basal
program
Confirm that correct basal program
is active.
Confirm that PDM time is set
correctly.
Consult your healthcare provider
about adjusting your basal pro-
grams or using a temporary (temp)
basal rate.
Incorrect bolus
timing
Check carb intake.
Take bolus with food.
Check blood glucose before giving
meal bolus; if necessary, adjust
bolus.
Consult healthcare provider for
guidance.
Bolus too small