EasyManua.ls Logo

Insulet OmniPod - Examples of Suggested Bolus Calculations

Insulet OmniPod
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Up/Down Controller User Info/Support Blood Glucose Records
Appendices
133
Examples of suggested bolus calculations
The examples below use these settings:
Target BG: 100 mg/dL
I/C ratio: 1:15
Correction factor (CF): 50
Duration of insulin action: 3 hrs
Example 1
Current BG level: 150 mg/dL
Carb intake: 0 grams of carbohydrate (carbs)
Insulin on board: 0 units
Correction bolus: 150 mg/dL (current) – 100 mg/dL (target) = 50
50 ÷ 50 (CF) = 1 unit
Meal bolus: 0 (carbs) ÷ 15 (I/C ratio) = 0 units
Suggested bolus dose = 1 (correction) + 0 (meal) 0 (IOB) = 1U
Example 2
Current BG level: 150 mg/dL
Carb intake: 45 carbs
Insulin on board: 0 units
Correction bolus: 150 mg/dL (current) – 100 mg/dL (target) = 50
50 ÷ 50 (CF) = 1 unit
Meal bolus: 45 (carbs) ÷ 15 (I/C ratio) = 3 units
Suggested bolus dose = 1 (correction) + 3 (meal) 0 (IOB) = 4U
Example 3
Current BG level: 100 mg/dL
Carb intake: 45 carbs
Insulin on board: 1 unit
Correction bolus: 100 mg/dL (current) – 100 mg/dL (target) = 0
0 ÷ 50 (CF) = 0 units
Meal bolus: 45 (carbs) ÷ 15 (I/C ratio) = 3 units
Suggested bolus dose = 0 (correction) + 3 (meal) 1 (IOB) = 2U
Example 4
Current BG level: 150 mg/dL
Carb intake: 45 carbs
Insulin on board: 1 unit
Correction bolus: 150 mg/dL (current) – 100 mg/dL (target) = 50
50 ÷ 50 (CF) = 1 unit
Meal bolus: 45 (carbs) ÷ 15 (I/C ratio) = 3 units
Suggested bolus dose = 1 (correction) + 3 (meal) 1 (IOB) = 3U