Up/Down Controller User Info/Support Blood Glucose Records
Checking Your Blood Glucose
7
73
The control solution test results should fall within the range
printed on the test strip vial label.
If control solution results are out of this range, REPEAT THE
TEST.
Out-of-range control solution results
Out-of-range test results may be caused by:
Expired or bad control solution
Expired or bad test strip
Error in performing test
Watered-down control solution
Code on test strip vial does not match code set in the PDM
Malfunction of the OmniPod System
Control solution test done outside 59º to 104ºF (15º to 40ºC)
If your control solution test results continue to fall outside the
range printed on the test strip vial:
1. The OmniPod System may not be working properly.
2. Do not use the System to test your blood glucose.
3. Call Customer Support, 800-591-3455, 24 hours/7 days
(from outside the United States, 781-457-5098).
Performing a Control Solution Test
1. Grasp the bottom of the test strip with the name “FreeStyle®”
facing up (Figure 7-3).
The control solution range is a target range for control
solution only. It is not a target range for your blood
glucose.
Figure 7-3
Top, to insert into meter
FreeStyle® test strip
Bottom, for blood sample
Edges of test strip