Checking Your Blood Glucose
72
When to perform a control solution test
You should perform a control solution test:
When you first get your OmniPod System, before doing a
blood glucose test
When you open and begin using a new vial of test strips
When you suspect that your meter or test strips are not work-
ing properly
When you think your test results are not accurate, or if your
test results are not consistent with how you feel
When you drop or damage your PDM or expose it to liquids
When your healthcare provider advises you to do so
Check that the meter and test strips are working
A control solution test follows the same procedure as a blood
glucose test, except that you use a sample of control solution
instead of a drop of blood. (See “Performing a Blood Glucose
Test” later in this chapter.)
When you perform a control solution test, if the reading is within
the control solution acceptable range, the meter is working
properly.
Control solution expiration date
FreeStyle® control solution is good for 3 months after opening
the bottle, or until the expiration date printed on the label,
whichever comes first. Count forward 3 months from the date
you open a new bottle of control solution. This is your discard
date. Write this date on the side of the control solution bottle. For
example, if you open the control solution on January 15, count
forward 3 months to April 15. This is the discard date to record
on the bottle, and the date to discard it and begin using a new
bottle.
Ensure accurate control solution results
Replace the cap on the control solution bottle immediately
after using it.
Do not add water or any liquid to control solution.
Control solution tests are specified to be accurate only when
performed between 59° and 104° F (15° and 40° C).
Use only FreeStyle® control solution with the
OmniPod System. Other brands of control solu-
tion may produce inaccurate results with this
System.
Do not use control solution past the expiration
date or you may get inaccurate results.
Results from FreeStyle® control solution tests do
not reflect your blood glucose level.