Up/Down Controller User Info/Support Blood Glucose Records
Errors, Advisories, and Hazard Alarms
10
117
PDM 2 sets of beeps every 5 minutes. Check the key described on the
screen. (In this example, the stuck key
is the Home button.) Press the key to
unstick it.
If this does not work, call Customer
Support at 800-591-3455 for assis-
tance (from outside the United States,
781-457-5098).
Pod 2 sets of beeps every 30 seconds for 5
minutes.
Escalates to Occusion detected hazard
alarm if not resolved within 5 minutes.
Check infusion site and pod for any
signs of occlusion.
Press OK to acknowledge.
Check blood glucose.
Alarm Source Beep Pattern Action
Stuck key detected
Pod occlusion advisory
Advisory Alarms