2600AS-901-01 Rev. B / September 2008 Return to Section Topics 19-51
Series 2600A System SourceMeter® Instruments Reference Manual Section 19: Remote Commands
display.trigger
Usage func = display.smuX.measure.func -- Reads function.
display.smuX.measure.func = func -- Writes function.
Set func to one of the following values:
0 or display.MEASURE_DCAMPS Selects current measure function.
1 or display.MEASURE_DCVOLTS Selects volts measure function.
2 or display.MEASURE_OHMS Selects ohms measure function.
3 or display.MEASURE_WATTS Selects power measure function.
Remarks • Selects the displayed measurement function: amps, volts, ohms, or watts.
• SMU A and SMU B can be set for different measurement functions.
Details See Measurement functions in Section 11.
Example Selects the current measure function for SMU A:
display.smua.measure.func = display.MEASURE_DCAMPS
display.trigger.clear
Function Clears the front panel trigger event detector.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage display.trigger.clear()
Remarks • The trigger event detector remembers if an event has been detected since the last
display.trigger.wait call. This function clears the trigger’s event detector and
discards the previous history of TRIG key presses.
• This attribute also clears the display.trigger.overrun attribute.
Also see display.trigger.wait, display.trigger.overrun
display.trigger.EVENT_ID
Attribute The trigger event number.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage event_id = display.trigger.EVENT_ID
event_id The trigger event number.
Remarks • Set the stimulus of any trigger event detector to the value of this constant to have it respond to
front panel trigger events.
display.trigger.overrun
Attribute The event detector overrun status.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage overrun = display.trigger.overrun
overrun The trigger overrun state.