Features and Functions 3
Keysight 33210A User’s Guide 109
Marker Frequency
If desired, you can set the frequency at which the signal on the front-panel
Sync connector goes to a logic low during the sweep. The Sync signal always
goes from low to high at the beginning of the sweep.
– Marker frequency: 1 mHz to 10 MHz (limited to 100 kHz for ramp waveforms).
The default is 500 Hz.
– When the sweep mode is enabled, the marker frequency must be between the
specified start frequency and stop frequency. If you attempt to set the marker
frequency to a frequency not in this range, the function generator will
automatically
set the marker frequency equal to the start frequency or stop
frequency
(whichever is closer).
– The Sync enable setting is overridden by enabling the marker used with the
sweep mode (see page 81). Therefore, when
the marker is enabled (and the
sweep mode is also enabled
), the Sync setting is ignored.
– Front-Panel Operation: After enabling sweeps, press the Marker softkey. Then
use the knob or numeric keypad to enter the desired marker frequency.
– Remote Interface Operation:
MARKer:FREQuency {<frequency>|MINimum|MAXimum}
MARKer {Off|On}