Remote Interface Reference 4
Keysight 33210A User’s Guide 223
Sweep Commands
FREQuency:STARt {<frequency>|MINimum|MAXimum}
FREQuency:STARt? [MINimum|MAXimum]
Set the start frequency (used in conjunction with the stop frequency).
Select from
1 mHz to 10 MHz
(limited to 100 kHz for a ramp waveform). The default is 100 Hz.
MIN = 1 mHz. MAX = 10 MHz. The :STAR? query returns the start frequency in
hertz.
–To sweep up in frequency, set the start frequency < stop frequency.
To sweep
down
in frequency, set the start frequency > stop frequency.
FREQuency:STOP {<frequency>|MINimum|MAXimum}
FREQuency:STOP? [MINimum|MAXimum]
Set the stop frequency (used in conjunction with the start frequency).
Select from
1 mHz to 10 MHz
(limited to 100 kHz for a ramp waveform). The default is 1 kHz.
MIN = 1 mHz. MAX = 10 MHz. The :STOP? query returns the stop frequency in
hertz.