Front-Panel Menu Operation 2
Keysight 33210A User’s Guide 59
To Configure the Remote Interface
The Keysight 33210A supports remote interface communication using a choice of
three interfaces: GPIB, USB, and LAN (LXI Class C compliant). All three interfaces
are “live” at power up. The following sections tell how to configure the remote
interface from the instrument front panel.
GPIB Configuration
You need only select a GPIB address.
1 Select the “I/O” menu.
Press and then press the
I/O
softkey.
2 Set the GPIB address.
Use the knob and cursor keys or the numeric keypad to select a GPIB address
in the range 0 through 30 (the factory default is “10”).
The GPIB address is shown on the front-panel display at power-on.
3 Exit the menu.
Press the DONE softkey.
NOTE
Two CDs, provided with your instrument, contain connectivity software to enable
communications over the remote interfaces. See “Connectivity Software and
Product CDs” on page 143, for further information on these CDs and the
software they contain.