4 Remote Interface Reference
180 Keysight 33210A User’s Guide
Using the APPLy Command
See also “Output Configuration” starting on page 68 in Chapter 3.
The APPLy command provides the most straightforward method to program the
function generator over the remote interface. You can select the function,
frequency, amplitude, and offset all in one command as shown in the syntax
statement below.
APPLy:<function> [<frequency> [,<amplitude> [,<offset>] ]]
For example, the following command string sent from your computer will output a
3 Vpp sine wave at 5 kHz with a -2.5 volt offset.
APPL:SIN 5 KHZ, 3.0 VPP, -2.5 V
The APPLy command performs the following operations:
Sets the trigger source to Immediate (equivalent to sending the TRIG:SOUR
IMM command).
Turns off any modulation, sweep, or burst mode currently enabled and places
the instrument in the continuous waveform mode.
Turns on the Output connector (OUTP ON command) but does not
change the
output termination setting (
OUTP:LOAD
command).
Overrides the voltage autorange setting
and automatically enables autoranging
(
VOLT:RANG:AUTO
command
).
For square waveforms, overrides the current duty cycle setting and
automatically selects 50% (FUNC:SQU:DCYC command).
For ramp waveforms, overrides the current symmetry setting and
automatically selects 100% (FUNC:RAMP:SYMM command).
The syntax statements for the APPLy command are shown on page 185.