5 Error Messages
324 Keysight 33210A User’s Guide
Calibration Errors
The following errors indicate failures that may occur during a calibration procedure
(s
ee Chapter 4 in the Keysight 33210A Service Guide).
701
Calibration error; security defeated by hardware jumper
The function generator’s calibration security feature has been disabled
by
temporarily shorting the two “CAL ENABLE” pads on the internal circuit board, as
described in the
Keysight 33210A
Service Guide
.
702
Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured. To
unsecure the instrument, use the CAL:SEC:STAT ON command with the correct
security code.
703
Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command.
706
Calibration error; provided value is out of range
The calibration value specified with the
CAL:VAL
command is out of range.
707
Calibration error; signal input is out of range
The internal analog-to-digital converter (ADC) has determined that the signal
applied to the rear-panel Modulation In connector is out of range.