3 Features and Functions
120 Keysight 33210A User’s Guide
Trigger Out Signal
A “trigger out” signal is provided on the rear-panel Trig Out connector (used with
sweep only). When enabled, a TTL-compatible pulse waveform
with either a rising
edge (default) or falling edge is output from the
Trig Out
connector
at the beginning
of the burst.
–When the Internal (immediate) trigger source is selected the function
generator outputs a square waveform with a 50% duty cycle from the Trig Out
connector at the beginning of the burst. The frequency of the waveform is
equal to the specified burst period.
–When the Manual trigger source is selected, the function generator outputs a
pulse (>1 µs pulse width) from the Trig Out connector at the beginning of each
burst.
–When the External trigger source is selected, the function generator
automatically disables the “trigger out” signal. The Trig Out connector cannot
be used for both operations at the same time (an externally-triggered
waveform uses the same connector to trigger the burst).
Front-Panel Operation
: After enabling bursts, press the
Trigger Setup
softkey.
Then select the desired edge by pressing the Trig Out softkey.
Remote Interface Operation:
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger {OFF|ON}