EasyManua.ls Logo

Keysight Technologies 33210A - Instrument Errors

Keysight Technologies 33210A
367 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Error Messages
320 Keysight 33210A User’s Guide
Instrument Errors
501 to 502
501: Cross-isolation UART framing error
502: Cross-isolation UART overrun error
These errors indicate an internal hardware failure.
The isolation between chassis
ground circuits and floating circuits is controlled by an optical isolation barrier and
a serial link.
580
Reference phase-locked loop is unlocked
The PHAS:UNL:ERR:STAT has been enabled (“on”) and the internal phase-locked
loop that controls the frequency is currently unlocked. This error is most likely to
occur when the external reference is out of lock range.