Remote Interface Reference 4
Keysight 33210A User’s Guide 279
Questionable Data Register Commands
See the table on page 274 for the register bit definitions.
STATus:QUEStionable:CONDition?
Query the condition register in this group. This is a read-only register and bits are
not cleared when you read the register. A query of this register returns a decimal
value which corresponds to the binary-weighted sum of all bits set in the register.
STATus:QUEStionable[:EVENt]?
Query the event register in this register group. This is a read-only register
. Once a bit
is set, it remains set until cleared by this command or *CLS (clear status)
command. A query of this register returns a decimal value
which corresponds to
the binary-weighted sum of all bits set in the register
.
STATus:QUEStionable:ENABle <enable value>
STATus:QUEStionable:ENABle?
Enable bits in the enable register in this register group. The selected bits are then
reported to the Status Byte. A *CLS (clear status) will not clear the enable register
but it does clear all bits in the event register. The STATus:PRESet command
clears all bits in the enable register. To enable bits in the enable register, you must
write a decimal value which corresponds to the binary-weighted sum of the bits
you wish to enable in the register.
The :ENAB? query returns a decimal value which corresponds to the
binary-weighted sum of all bits enabled by the STAT:QUES:ENAB command.