3 Features and Functions
80 Keysight 33210A User’s Guide
Remote Interface Operation:
OUTPut {OFF|ON}
The APPLy command overrides the current setting and automatically enables
the Output connector.
Waveform Polarity
In the normal mode (default), the waveform goes positive during the first part of
the cycle. In the inverted mode, the waveform goes negative during the first part
of the cycle.
As shown in the examples below, the waveform is inverted relative to the offset
voltage. Any offset voltage present will remain unchanged when the waveform
is inverted.
When a waveform is inverted, the Sync signal associated with the waveform is
not inverted.
Front-Panel Operation: Press and select the Output Setup softkey. Then
press the Normal softkey again to toggle between the “Normal” and “Invert”
selections.
Remote Interface Operation:
OUTPut:POLarity {NORMal|INVerted}
No Offset Voltage With Offset Voltage
Inverted
Normal
0V