Remote Interface Reference 4
Keysight 33210A User’s Guide 261
Interface Configuration Commands
See also “Remote Interface Configuration” on page 143 in Chapter 3.
Remote/Local Instrument State Commands:
The following commands set the remote/local state of the function generator over
the LAN interface from a Telnet or socket session. These commands provide
control analogous to the IEEE-488.2 commands such as GTL (Go To Local) over
the GPIB and USB interfaces.
SYSTem:LOCal
Sets the instrument state to local (the normal power-on default state). Removes
any annunciator and unlocks the front panel keyboard.
SYSTem:REMote
Sets the instrument state to remote. Displays the remote annunciator and locks
the keyboard (except the key).
SYSTem:RWLock
Sets the instrument state to remote with lock. Displays the rwl annunciator and
locks the keyboard (including the key).
SYSTem:COMMunicate:RLSTate {LOCal|REMote|RWLock}
This command performs the same functions as the previous three separate
commands.
LOCal - (The default). Sets the instrument state to local. Removes any
annunciator and unlocks the front panel keyboard.
REMote - Sets the instrument state to remote. Displays the remote
annunciator and locks the keyboard (except the key).
RWLock - Sets the instrument state to remote with lock. Displays the rwl
annunciator and locks the keyboard (including the key).